Sensors (Oct 2024)

Development and Validation of a Portable EIT System for Real-Time Respiratory Monitoring

  • Fabian Alvarado-Arriagada,
  • Bruno Fernández-Arroyo,
  • Samuel Rebolledo,
  • Esteban J. Pino

DOI
https://doi.org/10.3390/s24206642
Journal volume & issue
Vol. 24, no. 20
p. 6642

Abstract

Read online

This work contributes to the improvement of novel medical technologies for the prevention and treatment of diseases. Electrical impedance tomography (EIT) has gained attention as a valuable tool for non-invasive monitoring providing real-time insights. The purpose of this work is to develop and validate a novel portable EIT system with a small form factor for respiratory monitoring. The device uses a 16-electrode architecture with adjacent stimulation and measurement patterns, an integrated circuit current source and a single high-speed ADC operating with multiplexers to stimulate and measure across all electrodes. Tests were conducted on 25 healthy subjects who performed a pulmonary function test with a flowmeter while using the EIT device. The results showed a good performance of the device, which was able to recognize all respirations correctly, and from the EIT signals and images, correlations of 96.7% were obtained for instantaneous respiratory rate and 96.1% for tidal volume prediction. These results validate the preliminary technical feasibility of the EIT system and demonstrates its potential as a reliable tool for non-invasive respiratory assessment. The significance of this work lies in its potential to democratize advanced respiratory monitoring technologies, making them accessible to a wider population, including those in remote or underserved areas.