Sensors (Sep 2011)

Gamma-Ray Sterilization Effects in Silica Nanoparticles/γ-APTES Nanocomposite-Based pH-Sensitive Polysilicon Wire Sensors

  • Po-Yen Hsu,
  • Jing-Jenn Lin

DOI
https://doi.org/10.3390/s110908769
Journal volume & issue
Vol. 11, no. 9
pp. 8769 – 8781

Abstract

Read online

In this paper, we report the γ-ray sterilization effects in pH-sensitive polysilicon wire (PSW) sensors using a mixture of 3-aminopropyltriethoxysilane (γ-APTES) and polydimethylsiloxane (PDMS)-treated hydrophobic fumed silica nanoparticles (NPs) as a sensing membrane. pH analyses showed that the γ-ray irradiation-induced sensitivity degradation of the PSW pH sensor covered with γ-APTES/silica NPs nanocomposite (γ-APTES+NPs) could be restored to a condition even better than prior to γ-ray irradiation by 40-min of post-sterilization room-temperature UV annealing. We found that the trapping charges caused by γ-ray sterilization primarily concentrated in the native oxide layer for the pH sensor covered with γ-APTES, but accumulated in the γ-APTES+NPs layer for the γ-APTES+NPs-covered sensor. It is believed that mixing the PDMS-treated silica NPs into g-APTES provides many g-APTES/SiO2 interfaces for the accumulation of trapping charges and for post-sterilization UV oxidation, thus restoring γ-ray-induced sensor degradation. The PDMS-treated silica NPs not only enhance the sensitivity of the pH-sensitive PSW sensors but are also able to withstand the two-step sterilization resulting from γ-ray and UV irradiations. This investigation suggests γ-ray irradiation could be used as a highly-efficient sterilization method for γ-APTES-based pH-sensitive biosensors.