CVIR Endovascular (Sep 2020)
Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting
Abstract
Abstract Background Emergency carotid artery stenting (CAS) is a frequent endovascular procedure, especially in combination with intracranial thrombectomy. Balloon guide catheters are frequently used in these procedures. Our aim was to determine if mechanical aspiration through the working lumen of a balloon occlusion catheter during the steps of a carotid stenting procedure achieve flow rates that may lead to internal carotid artery (ICA) flow reversal which consecutively may prevent distal embolism. Methods Aspiration experiments were conducted using a commercially available aspiration pump. Aspiration flow rates/min with 6 different types of carotid stents inserted into a balloon guide catheter were measured. Measurements were repeated three times with increasing pressure in the phantom. To determine if the achieved aspiration flow rates were similar to physiologic values, flow rates in the ICA and external carotid artery (ECA) in 10 healthy volunteers were measured using 4D-flow MRI. Results Aspiration flow rates ranged from 25 to 82 mL/min depending on the stent model. The pressure in the phantom had a significant influence on the aspiration volume. Mean blood flow volumes in volunteers were 210 mL/min in the ICA and 101 mL/min in the ECA. Conclusions Based on the results of this study, flow reversal in the ICA during common carotid artery occlusion is most likely achieved with the smallest diameter stent sheath and the stent model with the shortest outer stent sheath maximum diameter. This implies that embolic protection during emergency CAS through aspiration is most effective with these models.