Journal of Ophthalmic & Vision Research (Nov 2023)

Agreement Between Swept-source Optical Coherence Tomography and Rotating Scheimpflug Camera in Measurement of Corneal Parameters in Normal and Keratoconic Eyes

  • Hamed Ghassemi,
  • Mehran Zarei-Ghanavati,
  • Mina Khastavan,
  • Mehrnaz Atighehchian,
  • Abbas Azimi Khorasani,
  • Golshan Latifi

DOI
https://doi.org/10.18502/jovr.v18i4.14547
Journal volume & issue
Vol. 18, no. 4
pp. 386 – 395

Abstract

Read online

Abstract Purpose: This study aimed to assess the agreement between topographic indices of healthy subjects and keratoconus (KCN) patients using a swept-source optical coherence tomography (SS-OCT CASIA2) versus a Scheimpflug camera (Pentacam). Methods: 40 eyes of 23 patients with KCN and 40 eyes of 20 healthy subjects were included and evaluated with the CASIA2, followed by the Pentacam. Two consecutive modalities were obtained for one eye of each patient. Corneal parameters, including anterior keratometry at steep (Ks) and flat meridians (Kf), anterior astigmatism, anterior and posterior corneal elevation values, thinnest corneal thickness, and apex corneal thickness, were evaluated. Results: CASIA2 and Pentacam showed perfect agreement (95% limits of agreement (LoA): -0.22 to 0.68, 95% LoA: -1.5 to 1.44 D) and good correlation (Intraclass correlation (ICC):0.986, ICC:0.987; P < 0.01) for anterior (Ks) in normal and ectatic corneas, respectively. The cylinder amount had moderate agreement and correlation (95% LoA: -0.55 to 0.47D, ICC: 0.797, P < 0.01) in normal, and moderate to strong agreement and correlation (95% LoA: -1.57 to 0.87D, ICC=0.911, P < 0.01) in Keratoconic eyes. There was a fair agreement for anterior and posterior corneal elevation values in normal subjects (95% LoA: -3.09 to 4.59, 95% LoA: -6.91 to 7.31D). The thinnest corneal thickness amount had an excellent agreement in normal and KCN patients (ICC: 0.983, 0.953; respectively). Conclusions: Although the devices had different mean indices values, they had a good agreement based on the Bland–Altman plots. Since Pentacam is accepted as the standard tool for diagnosing ectatic cornea, pentacam CASIA2 is also helpful for early diagnosis of KCN.

Keywords