PLoS ONE (Jan 2016)
Non-Invasive Evaluation of Heart Function with Four-Dimensional Echocardiography.
Abstract
BACKGROUND:The aim of this study is to assess the accuracy and feasibility of left ventricular systolic function determined by four-dimensional echocardiography (4DE). METHODS:Latex balloons were sewn into the left ventricle (LV) of 20 freshly harvested pig hearts which were then passively driven by a pulsatile pump apparatus. Global longitudinal strain (GLS), global circumferential strain (GCS), global area strain (GAS) and left ventricular ejection fraction (LVEF) derived from 4DEand two-dimensional echocardiography (2DE)-derived LVEF were quantified at different stroke volumes (SV) 30-70 ml and correlated with sonomicrometry data. RESULTS:In all comparisons, GLS, GCS, GAS, 2DE-LVEF, and 4DE-LVEF demonstrated strong correlations with sonomicrometry data (r = 0.77, r = 0.89, r = 0.79, r = 0.93, r = 0.96, all P <0.001). Bland-Altman analyses showed slight overestimations of echo-derived GLS, GCS, 2DE-LVEF and 3DE-LVEF over sonomicrometry values (bias = 2.88, bias = 3.99, bias = 3.37, bias = 2.78, respectively). Furthermore, there is better agreement between GCS, 4D LVEF and sonomicrometry values compared with GLS and 2D LVEF. CONCLUSION:Four-dimensional echocardiography accurately assesses LV function. GCS derived by 4DE is a potential alternative parameter to quantify LV systolic function.