Pharmaceutics (Jan 2024)

Mechanical Characterization of Dissolving Microneedles: Factors Affecting Physical Strength of Needles

  • Daisuke Ando,
  • Megumi Miyatsuji,
  • Hideyuki Sakoda,
  • Eiichi Yamamoto,
  • Tamaki Miyazaki,
  • Tatsuo Koide,
  • Yoji Sato,
  • Ken-ichi Izutsu

DOI
https://doi.org/10.3390/pharmaceutics16020200
Journal volume & issue
Vol. 16, no. 2
p. 200

Abstract

Read online

Dissolving microneedles (MNs) are novel transdermal drug delivery systems that can be painlessly self-administered. This study investigated the effects of experimental conditions on the mechanical characterization of dissolving MNs for quality evaluation. Micromolding was used to fabricate polyvinyl alcohol (PVA)-based dissolving MN patches with eight different cone-shaped geometries. Axial force mechanical characterization test conditions, in terms of compression speed and the number of compression needles per test, significantly affected the needle fracture force of dissolving MNs. Characterization using selected test conditions clearly showed differences in the needle fracture force of dissolving MNs prepared under various conditions. PVA-based MNs were divided into two groups that showed buckling and unbuckling deformation, which occurred at aspect ratios (needle height/base diameter) of 2.8 and 1.8, respectively. The needle fracture force of PVA-based MNs was negatively correlated with an increase in the needle’s aspect ratio. Higher residual water or higher loading of lidocaine hydrochloride significantly decreased the needle fracture force. Therefore, setting appropriate methods and parameters for characterizing the mechanical properties of dissolving MNs should contribute to the development and supply of appropriate products.

Keywords