Frontiers in Medical Technology (May 2024)

Experimental validation of auxetic stent designs: three-point bending of 3D printed Titanium prototypes

  • Rahul Vellaparambil,
  • Rahul Vellaparambil,
  • Woo-Suck Han,
  • Pierluigi Di Giovanni,
  • Stéphane Avril

DOI
https://doi.org/10.3389/fmedt.2024.1388207
Journal volume & issue
Vol. 6

Abstract

Read online

IntroductionNumerical simulations have demonstrated the superior bending flexibility of auxetic stents compared to conventional stent designs for endovascular procedures. However, conventional stent manufacturing techniques struggle to produce complex auxetic stent designs, fueling the adoption of additive manufacturing techniques.MethodsIn this study, we employed DMLS additive manufacturing to create Titanium Ti64 alloy stent prototypes based on auxetic stent designs investigated in a previous study. These prototypes were then subjected to experimental three-point bending tests.ResultThe experimental results were replicated using a finite element model, which showed remarkable accuracy in predicting the bending flexibility of four auxetic stents and two conventional stents.DiscussionAlthough this validation study demonstrates the promising potential of DMLS and other additive manufacturing methods for fabricating auxetic stents, further optimization of current stent design limitations and the incorporation of post-processing techniques are essential to enhance the reliability of these additive manufacturing processes.

Keywords