Physics and Imaging in Radiation Oncology (Jul 2020)
Audit feasibility for geometric distortion in magnetic resonance imaging for radiotherapy
Abstract
Background and purpose: Magnetic Resonance Imaging (MRI) is increasingly being used in radiotherapy (RT). However, geometric distortions are a known challenge of using MRI in RT. The aim of this study was to demonstrate feasibility of a national audit of MRI geometric distortions. This was achieved by assessing large field of view (FOV) MRI distortions on a number of scanners used clinically for RT. Materials and methods: MRI scans of a large FOV MRI geometric distortion phantom were acquired on 11 MRI scanners that are used clinically for RT in the UK. The mean and maximum distortions and variance between scanners were reported at different distances from the isocentre. Results: For a small FOV representing a brain (100–150 mm from isocentre) all distortions were 10 mm in some cases. The variance between scanners was low and was found to increase with distance from isocentre. Conclusions: This study demonstrated feasibility of the technique to be repeated in a country wide geometric distortion audit of all MRI scanners used clinically for RT. Recommendations were made for performing such an audit and how to derive acceptable limits of distortion in such an audit.