Abstract
A new voltammetric sensor based on an aryl amide type podand, 1,8-bis(o-amidophenoxy)-3,6-dioxaoctane, (AAP) modified glassy carbon electrode, was described for the determination of trace level of mercury (II) ion by cyclic voltammetry (CV) and differential pulse voltammetry (DPV). A well-defined anodic peak corresponding to the oxidation of mercury on proposed electrode was obtained at 0.2 V versus Ag/AgCl reference electrode. The effect of experimental parameters on differential voltammetric peak currents was investigated in acetate buffer solution of pH 7.0 containing 1 × 10−1 mol L−1 NaCl. Mercury (II) ion was preconcentrated at the modified electrode by forming complex with AAP under proper conditions and then reduced on the surface of the electrode. Interferences of Cu2+, Pb2+, Fe3+, Cd2+, and Zn2+ ions were also studied at two different concentration ratios with respect to mercury (II) ions. The modified electrode was applied to the determination of mercury (II) ions in seawater sample.