BMJ Open (Apr 2016)

Retrospective cohort study of the performance of the Pinnacle metal on metal (MoM) total hip replacement: a single-centre investigation in combination with the findings of a national retrieval centre

  • Thomas Joyce,
  • David John Langton,
  • Raghavendra Prasad Sidaginamale,
  • Peter Avery,
  • Sue Waller,
  • Ghanshyabhai Tank,
  • James Lord,
  • Nick Cooke,
  • Raj Logishetty,
  • Antoni Viraf Francis Nargol

DOI
https://doi.org/10.1136/bmjopen-2015-007847
Journal volume & issue
Vol. 6, no. 4

Abstract

Read online

Objectives To determine risk factors for revision in patients implanted with a commonly used metal on metal (MoM) hip replacement.Design Retrospective cohort study in combination with a prospective national retrieval study (Northern Retrieval Registry (NRR)).Setting Combined orthopaedic unit in combination with the NRR.Participants All patients implanted with a DePuy Pinnacle MoM hip prostheses by the 2 senior authors were invited to attend for a review which included clinical examination, blood metal ion measurements, radiographs and targeted imaging. Explanted components underwent wear analysis using validated methodology and these results were compared with those obtained from the NRR.Results 489 MoM Pinnacle hips were implanted into 434 patients (243 females and 191 males). Of these, 352 patients attended the MoM recall clinics. 64 patients had died during the study period. For the purposes of survival analysis, non-attendees were assumed to have well-functioning prostheses. The mean follow-up of the cohort as a whole was 89 months. 71 hips were revised. Prosthetic survival for the whole cohort was 83.6% (79.9–87.3) at 9 years. The majority of explanted devices exhibited signs of taper junction failure. Risk factors for revision were bilateral MoM prostheses, smaller Pinnacle liners, and implantation in 2006 and later years. A significant number of devices were found to be manufactured out of their specifications. This was confirmed with analysis of the wider data set from the NRR.Conclusions This device was found to have an unacceptably high revision rate. Bilateral prostheses, those implanted into female patients and devices implanted in later years were found to be at greater risk. A significant number of explanted components were found to be manufactured with bearing diameters outside of the manufacturer's stated tolerances. Our findings highlight the clinical importance of hitherto unrecognised variations in device production.