Data in Brief (Feb 2018)

Data from the analytical performance of the Abaxis Piccolo Xpress point of care analyzer in whole blood, serum, and plasma

  • Kazunori Murata,
  • Laurel Glaser,
  • Mary Nardiello,
  • Lakshmi V. Ramanathan,
  • Dean C. Carlow

Journal volume & issue
Vol. 16
pp. 81 – 89

Abstract

Read online

The objective of this study was to examine the analytical performance of 14 comprehensive metabolic panel analytes on the Abaxis Piccolo Xpress® Point of Care analyzer in serum, plasma, and whole blood. A method comparison was performed on all three specimen types intended for use on the Piccolo Xpress®: serum, heparinized plasma, and whole blood. This data is also presented in Murata et al. (2015) [1]. This article includes the actual Bland-Altman bias plots of the difference in results obtained for analytes in the comprehensive metabolic panel from the Abaxis Piccolo Xpress and the comparison instrument, the Ortho Vitros. Keywords: Clinical chemistry, Point-of care testing