Scientific Reports (Jul 2024)
Green method for 17-hydroxyprogesterone extraction and determination using PDMS stir bar sorptive extraction coupled with HPLC: optimization by response surface methodology
Abstract
Abstract Quantifying small amounts of the 17-hydroxyprogesterone in various matrix is crucial for different purposes. In this study, a commercial polydimethylsiloxane stir bar was used to extract hormone from water and urine samples. Analysis was performed by high-performance liquid chromatography using a UV detector. The response surface methodology was used to optimize the desorption and extraction steps, with predicted optimal point relative errors of 1.25% and 6.40%, respectively. The optimized method was validated with a linear range of 1.21–1000.00 for aqueous and 2.43–2000.00 ng mL–1 for urine samples. The coefficient of determination was 0.9998 and 0.9967, and the detection limit of the proposed method was obtained to be 0.40 and 0.80 ng mL–1 for aqueous and urine samples, respectively. The recovery percentage and relative standard deviation within a day and between three days after the addition of three different concentration levels of the standard to the control sample were 87–103% and 0.4–3.6% for aqueous and 87.5–101% and 0.1–5.2% for urine samples, respectively. The results show that the proposed method can be appropriate and cost-effective for extracting and analyzing this hormone. In addition, using three different tools, the greenness of the proposed method was proven.
Keywords