PLoS ONE (Jan 2021)

Evaluation of two automated low-cost RNA extraction protocols for SARS-CoV-2 detection.

  • Fernando Lázaro-Perona,
  • Carlos Rodriguez-Antolín,
  • Marina Alguacil-Guillén,
  • Almudena Gutiérrez-Arroyo,
  • Jesús Mingorance,
  • Julio García-Rodriguez,
  • SARS-CoV-2 Working Group

DOI
https://doi.org/10.1371/journal.pone.0246302
Journal volume & issue
Vol. 16, no. 2
p. e0246302

Abstract

Read online

BackgroundTwo automatable in-house protocols for high-troughput RNA extraction from nasopharyngeal swabs for SARS-CoV-2 detection have been evaluated.MethodsOne hundred forty one SARS-CoV-2 positive samples were collected during a period of 10-days. In-house protocols were based on extraction with magnetic beads and designed to be used with either the Opentrons OT-2 (OT-2in-house) liquid handling robot or the MagMAXTM Express-96 system (MMin-house). Both protocols were tested in parallel with a commercial kit that uses the MagMAXTM system (MMkit). Nucleic acid extraction efficiencies were calculated from a SARS-CoV-2 DNA positive control.ResultsNo significant differences were found between both in-house protocols and the commercial kit in their performance to detect positive samples. The MMkit was the most efficient although the MMin-house presented, in average, lower Cts than the other two. In-house protocols allowed to save between 350€ and 400€ for every 96 extracted samples compared to the commercial kit.ConclusionThe protocols described harness the use of easily available reagents and an open-source liquid handling system and are suitable for SARS-CoV-2 detection in high throughput facilities.