Scientific Reports (Feb 2024)

Application of anatomy-based spacing of electrode contacts for achieving a uniform semitonal resolution: A novel concept in cochlear implant electrode design

  • Isra Ali Aljazeeri,
  • Abdulrahman Hagr

DOI
https://doi.org/10.1038/s41598-024-53070-8
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Using anatomy-based fitting, we can determine the place-specific map with individualized center frequencies for each electrode contact that is a closer match to the natural pitch-place of the cochlea. The primary objective of this study is to evaluate the tonal presentation across the electrode array and to calculate the semitone difference between each adjacent pair of contacts according to their anatomy-based map. The secondary objective is to determine the distancing of the contacts that would result in an equal semitone difference with a uniform tonal presentation. A total of 167 ears were included in this retrospective study. The frequencies across the electrode arrays were found to be unequally presented. The semitonal condensations were higher in the apical inter-contact spaces compared to the basal inter-contact spaces, being 3.0–2.3 semitones/mm (Kruskal Wallis test, p < 0.000). The anatomy-based spacing of the electrode contacts was larger in the basal inter-contact spaces compared to the apical inter-contact spaces, ranging from 1.92 to 1.48 mm. In conclusion, the current electrode designs do not have uniform tonal representation throughout the electrode array. There is a more condensed tonal presentation in the apical electrodes than in the basal electrodes, resulting in a lower tonal resolution in the apical region.