Iranian Journal of Medical Physics (May 2020)

Comparing IDREAM as an Iterative Reconstruction Algorithm against In Filtered Back Projection in Computed Tomography

  • Eslam Maamoun

DOI
https://doi.org/10.22038/ijmp.2019.41872.1613
Journal volume & issue
Vol. 17, no. 3
pp. 170 – 174

Abstract

Read online

Introduction: Recent studies of Computed Tomography (CT) conducted on patient dose reduction have recommended using an iterative reconstruction algorithm and mA (mili-Ampere) dose modulation. The current study aimed to evaluate Iterative Dose Reduction Algorithm (IDREAM) as an iterative reconstruction algorithm. Material and Methods: Two CT protocols (i.e., A: 120 KV /150 mA, FBP; B: 120KV/ (20-150) mAs, IDREAM) to scan water and acrylic phantoms. A number of 40 patients were assigned to two CT protocols (C: n=20, 120KV/160 ±10 mAs, FBP and D: n=20, 120 KV/ (30-150 mAs, IDREAM), the two groups (C and D) were then referred to abdomen and pelvis CT scan (Sinovision, insitum 16) with contrast. Image quality parameters, dose calculations were measured for all groups (i.e., A, B, C, and D). Results: Group B had a highly significant SNR with less significant noise (P0.05) in acrylic phantom, as compared to group A. CTDIvol (A: 13.94 mGy ; B: 6.91 mGy , PConclusion: The results of the present study were indicative of the feasibility of IDREAMas an iterative reconstruction algorithm.