Scientific Reports (Oct 2024)

Lateral flow assay sensitivity and signal enhancement via laser µ-machined constrains in nitrocellulose membrane

  • Gazy Khatmi,
  • Tomas Klinavičius,
  • Martynas Simanavičius,
  • Laimis Silimavičius,
  • Asta Tamulevičienė,
  • Agnė Rimkutė,
  • Indrė Kučinskaitė-Kodzė,
  • Gintautas Gylys,
  • Tomas Tamulevičius

DOI
https://doi.org/10.1038/s41598-024-74407-3
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 14

Abstract

Read online

Abstract Lateral flow assay (LFA) is a handful diagnostic technology that can identify severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and other common respiratory viruses in one strip, which can be tested at the point-of-care without the need for equipment or skilled personnel outside the laboratory. Although its simplicity and practicality make it an appealing solution, it remains a grand challenge to substantially enhance the colorimetric LFA sensitivity. In this work, we present a straightforward approach to enhance the sensitivity of LFA by imposing the flow constraints in nitrocellulose (NC) membranes via a number of vertical femtosecond laser micromachined microchannels which is important for prolonged specific binding interactions. Porous NC membrane surfaces were structured with different widths and densities µ-channels employing a second harmonic of the Yb:KGW femtosecond laser and sample XYZ translation over a microscope objective-focused laser beam. The influence of the microchannel parameters on the vertical wicking speed was evaluated from the video recordings. The obtained results indicated that µ-channel length, width, and density in NC membranes controllably increased the immunological reaction time between the analyte and the labeled antibody by 950%. Image analysis of the colorimetric indicators confirmed that the flow rate delaying strategy enhanced the signal sensitives by 40% compared with pristine NC LFA.

Keywords