Sensors (Dec 2021)

Integrated Relaxation Pressure Classification and Probe Positioning Failure Detection in High-Resolution Esophageal Manometry Using Machine Learning

  • Zoltan Czako,
  • Teodora Surdea-Blaga,
  • Gheorghe Sebestyen,
  • Anca Hangan,
  • Dan Lucian Dumitrascu,
  • Liliana David,
  • Giuseppe Chiarioni,
  • Edoardo Savarino,
  • Stefan Lucian Popa

DOI
https://doi.org/10.3390/s22010253
Journal volume & issue
Vol. 22, no. 1
p. 253

Abstract

Read online

High-resolution esophageal manometry is used for the study of esophageal motility disorders, with the help of catheters with up to 36 sensors. Color pressure topography plots are generated and analyzed and using the Chicago algorithm a final diagnosis is established. One of the main parameters in this algorithm is integrated relaxation pressure (IRP). The procedure is time consuming. Our aim was to firstly develop a machine learning based solution to detect probe positioning failure and to create a classifier to automatically determine whether the IRP is in the normal range or higher than the cut-off, based solely on the raw images. The first step was the preprocessing of the images, by finding the region of interest—the exact moment of swallowing. Afterwards, the images were resized and rescaled, so they could be used as input for deep learning models. We used the InceptionV3 deep learning model to classify the images as correct or failure in catheter positioning and to determine the exact class of the IRP. The accuracy of the trained convolutional neural networks was above 90% for both problems. This work is just the first step in fully automating the Chicago Classification, reducing human intervention.

Keywords