PeerJ (Jun 2024)

Recognition of apnea and hypopnea by non-contact optical fiber mattress and its application in the diagnosis of obstructive sleep apnea hypopnea syndrome: a retrospective study

  • Ling Zhou,
  • Xiaoyan Zhu,
  • Lu Liu,
  • Lingling Wang,
  • Pengdou Zheng,
  • Fengqin Zhang,
  • Zhenyu Mao,
  • Guoliang Huang,
  • Songlin Cheng,
  • Huiguo Liu,
  • Wei Liu

DOI
https://doi.org/10.7717/peerj.17570
Journal volume & issue
Vol. 12
p. e17570

Abstract

Read online Read online

Objectives This study sought to evaluate the diagnostic value of a non-contact optical fiber mattress for apnea and hypopnea and compare it with traditional polysomnography (PSG) in adult obstructive sleep apnea hypopnea syndrome (OSAHS). Methods To determine the value of a non-contact optical fiber mattress for apnea and hypopnea, six healthy people and six OSAHS patients were selected from Tongji Hospital to design a program to identify apnea or hypopnea. A total of 108 patients who received polysomnography for drowsiness, snoring or other suspected OSAHS symptoms. All 108 patients were monitored with both the non-contact optical fiber mattress and PSG were collected. Results Six healthy controls and six patients with OSAHS were included. The mean apnea of the six healthy controls was 1.22 times/h, and the mean hypopnea of the six healthy controls was 2 times/h. Of the six patients with OSAHS, the mean apnea was 12.63 times/h, and the mean hypopnea was 19.25 times/h. The non-contact optical fiber mattress results showed that the mean apnea of the control group was 3.17 times/h and the mean hypopnea of the control group was 3.83 times/h, while the mean apnea of the OSAHS group was 11.95 times/h and the mean hypopnea of the OSAHS group was 17.77 times/h. The apnea index of the non-contact optical fiber mattress was positively correlated with the apnea index of the PSG (P < 0.05, r = 0.835), and the hypopnea index of the non-contact optical fiber mattress was also positively correlated with the hypopnea index of the PSG (P < 0.05, r = 0.959). The non-contact optical fiber mattress had high accuracy (area under curve, AUC = 0.889), specificity (83.4%) and sensitivity (83.3%) for the diagnosis of apnea. The non-contact fiber-optic mattress also had high accuracy (AUC = 0.944), specificity (83.4%) and sensitivity (100%) for the diagnosis of hypopnea. Among the 108 patients enrolled, there was no significant difference between the non-contact optical fiber mattress and the polysomnography monitor in total recording time, apnea hypopnea index (AHI), average heart rate, tachycardia index, bradycardia index, longest time of apnea, average time of apnea, longest time of hypopnea, average time of hypopnea, percentage of total apnea time in total sleep time and percentage of total hypopnea time in total sleep time. The AHI value of the non-contact optical fiber mattress was positively correlated with the AHI value of the PSG (P < 0.05, r = 0.713). The specificity and sensitivity of the non-contact optical fiber mattress AHI in the diagnosis of OSAHS were 95% and 93%, with a high OSAHS diagnostic accuracy (AUC = 0.984). Conclusion The efficacy of the non-contact optical fiber mattress for OSAHS monitoring was not significantly different than PSG monitoring. The specificity of the non-contact optical mattress for diagnosing OSAHS was 95% and its sensitivity was 93%, with a high OSAHS diagnostic accuracy.

Keywords