Journal of Arrhythmia (Dec 2017)
Pacing failure caused by automatic pacing threshold adjustment system
Abstract
Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient. Keywords: Automatic pacing threshold adjustment algorithm, Pacing failure