Saudi Journal of Anaesthesia (Jan 2017)

Ultrasound-guided probe-generated artifacts stimulating ventricular tachycardia: A rare phenomenon

  • Rafat Shamim,
  • Rudrashish Haldar,
  • Ashutosh Kaushal

DOI
https://doi.org/10.4103/1658-354X.197340
Journal volume & issue
Vol. 11, no. 1
pp. 93 – 95

Abstract

Read online

Electrocardiographic (ECG) artifacts may arise due to interference, faulty earthing, and current leakages in biomedical equipment which might create clinical dilemmas in the perioperative settings. Piezoelectric signals generated by ultrasonography probe are another uncommon source which might be sensed by the ECG electrodes and produce tracings similar to pathological arrhythmias triggering false alarms and avoidable therapies. Anesthesiologists should be familiar with these uncommon sources which might produce these artifacts and they should be identified swiftly.