Scientific Reports (Aug 2021)

Phase stable swept-source optical coherence tomography with active mode-locking laser for contrast enhancements of retinal angiography

  • Kwan Seob Park,
  • Eunwoo Park,
  • Hwidon Lee,
  • Hyun-Ji Lee,
  • Sang-Won Lee,
  • Tae Joong Eom

DOI
https://doi.org/10.1038/s41598-021-95982-9
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Swept-source optical coherence tomography (SS-OCT) is an attractive high-speed imaging technique for retinal angiography. However, conventional swept lasers vary the cavity length of the laser mechanically to tune the output wavelength. This causes sweep-timing jitter and hence low phase stability in OCT angiography. Here, we improve an earlier phase-stabilized, akinetic, SS-OCT angiography (OCTA) method by introducing coherent averaging. We develop an active mode-locking (AML) laser as a high phase-stable akinetic swept source for the OCTA system. The phase stability of the improved system was analyzed, and the effects of coherent averaging were validated using a retina phantom. The effectiveness of the coherent averaging method was further confirmed by comparing coherently and conventionally averaged en face images of human retinal vasculature for their contrast-to-noise ratio, signal-to-noise ratio, and vasculature connectivity. The contrast-to-noise ratio was approximately 1.3 times larger when applying the coherent averaging method in the human retinal experiment. Our coherent averaging method with the high phase-stability AML laser source for OCTA provides a valuable tool for studying healthy and diseased retinas.