Journal of Personalized Medicine (May 2024)

An Intraoperative Method to Minimize Leg Length Discrepancy in Anterior Minimally Invasive Total Hip Arthroplasty—A Prospective Study

  • Mauro Girolami,
  • Roberto Bevoni,
  • Elena Artioli,
  • Renata Beluzzi,
  • Cosimo Vasco,
  • Silvio Caravelli,
  • Annalisa Baiardi,
  • Massimiliano Mosca

DOI
https://doi.org/10.3390/jpm14060573
Journal volume & issue
Vol. 14, no. 6
p. 573

Abstract

Read online

While several intraoperative devices have been described in the literature for assessing leg length discrepancy (LLD), none have been utilized during total hip arthroplasty (THA) performed via the Anterior Minimally Invasive Surgery (AMIS) approach. The aim of this prospective study was to evaluate the efficacy and accuracy of a compass device in assessing leg length during THA performed using the AMIS technique. A prospective study was conducted involving 35 patients who consecutively underwent unilateral primary THA using the AMIS technique at our department from September 2017 to December 2018. LLD was measured by comparing preoperative and postoperative anteroposterior radiographs of the pelvis, independently assessed by two observers. The mean preoperative LLD was 3.6 (SD 3.9, range, 0.2–19.3) mm. The mean postoperative LLD was 2.5 (SD 3.0, range, 0–12.2) mm. A postoperative LLD of less than 5 mm was observed in 88.2% of cases, with 94.1% having values less than 10 mm. In conclusion, the compass device emerged as a valuable tool for ensuring precise limb length control in THA with the AMIS approach, offering both efficiency and cost-effectiveness in clinical practice.