Sensors (Jun 2022)

Path-Planning System for Radioisotope Identification Devices Using 4π Gamma Imaging Based on Random Forest Analysis

  • Hideki Tomita,
  • Shintaro Hara,
  • Atsushi Mukai,
  • Keita Yamagishi,
  • Hidetake Ebi,
  • Kenji Shimazoe,
  • Yusuke Tamura,
  • Hanwool Woo,
  • Hiroyuki Takahashi,
  • Hajime Asama,
  • Fumihiko Ishida,
  • Eiji Takada,
  • Jun Kawarabayashi,
  • Kosuke Tanabe,
  • Kei Kamada

DOI
https://doi.org/10.3390/s22124325
Journal volume & issue
Vol. 22, no. 12
p. 4325

Abstract

Read online

We developed a path-planning system for radiation source identification devices using 4π gamma imaging. The estimated source location and activity were calculated by an integrated simulation model by using 4π gamma images at multiple measurement positions. Using these calculated values, a prediction model to estimate the probability of identification at the next measurement position was created by via random forest analysis. The path-planning system based on the prediction model was verified by integrated simulation and experiment for a 137Cs point source. The results showed that 137Cs point sources were identified using the few measurement positions suggested by the path-planning system.

Keywords