Uncertainty evaluation method for axi-symmetric measurement machines

  • Muelaner Jody Emlyn,
  • Keogh Patrick S.

DOI
https://doi.org/10.1051/ijmqe/2016011
Journal volume & issue
Vol. 7, no. 2
p. 206

Abstract

Read online

This paper describes a method of uncertainty evaluation for axi-symmetric measurement machines. Specialized measuring machines for the inspection of axisymmetric components enable the measurement of properties such as roundness (radial runout), axial runout and coning. These machines typically consist of a rotary table and a number of contact measurement probes located on slideways. Sources of uncertainty include the probe calibration process, probe repeatability, probe alignment, geometric errors in the rotary table, the dimensional stability of the structure holding the probes and form errors in the reference hemisphere which is used to calibrate the system. The generic method is described and an evaluation of an industrial machine is described as a worked example. Expanded uncertainties, at 95% confidence, were then calculated for the measurement of; radial runout (1.2 μm with a plunger probe or 1.7 μm with a lever probe); axial runout (1.2 μm with a plunger probe or 1.5 μm with a lever probe); and coning/swash (0.44 arcseconds with a plunger probe or 0.60 arcseconds with a lever probe).

Keywords