Separations (Oct 2021)

Determination of Fosetyl-Aluminum in Wheat Flour with Extract-Dilute-Shoot Procedure and Hydrophilic Interaction Liquid Chromatography Tandem Mass Spectrometry

  • Xianjiang Li,
  • Sheng Wang,
  • Zhen Guo,
  • Xiuqin Li,
  • Qinghe Zhang,
  • Hongmei Li

DOI
https://doi.org/10.3390/separations8110197
Journal volume & issue
Vol. 8, no. 11
p. 197

Abstract

Read online

Fosetyl-aluminum is a widely used ionic fungicide. This pesticide is not amenable to the common multi-residue sample preparation methods. Herein, this paper describes a novel method for the simple and sensitive determination of fosetyl-aluminum residue in wheat flour. The sample preparation method involved extraction with water under ultrasonication and subsequent dilution with six-fold acetonitrile. The fosetyl-aluminum concentration was determined by hydrophilic interaction liquid chromatography tandem mass spectrometry. The limit of detection and quantification were only 5 and 10 ng/g, respectively, which meet the requirement of the current European legislation. Matrix-matched linearity (r2 = 0.9999) was established in the range of 10–2000 ng/g. Satisfactory recoveries were achieved in the range of 95.6% to 105.2% for three levels of spiked samples (10, 50, and 100 ng/g). Finally, the method was applied to analyzing 75 wheat flour samples produced in four provinces in China. Two samples were positive with concentrations over the limit of detection. This is the first method focusing on fosetyl-aluminum determination in wheat flour with an extract-dilute-shoot strategy and is very promising for the routine quality control of fosetyl-aluminum in similar cereal matrices.