Practical Laboratory Medicine (Jul 2018)
Assessment of a flow cytometry technique for studying signaling pathways in platelets: Monitoring of VASP phosphorylation in clinical samples
Abstract
A recently released kit (PerFix EXPOSE) was reported to improve the measurement of the degree of phosphorylation of proteins in leukocytes by flow cytometry. We tested its adaptation for platelets to monitor vasodilator-stimulated phosphoprotein (VASP) phosphorylation, which is the basis of a currently used test for the assessment of the pharmacological response to P2Y12 antagonists (PLT VASP/P2Y12). The PerFix EXPOSE kit was compared to the PLT VASP/P2Y12 kit by using blood samples drawn at 24 h post clopidogrel dose from 19 patients hospitalized for a non-cardio-embolic ischemic stroke and treated with clopidogrel monotherapy for at least five days in an observational study. The platelet PerFix method was based on adaptation of the volume of the sample, the centrifugation speed and the incubation temperature. Poor agreement between prevention by adenosine diphosphate (ADP) of PGE1-induced cAMP-mediated VASP phosphorylation and ADP induced aggregation assessed by Light Transmittance Aggregometry was found. We found a significant correlation between the PLT VASP/P2Y12 kit and the PerFix EXPOSE kit. The PerFix EXPOSE kit may also be helpful to monitor adverse effects of second-generation tyrosine kinase inhibitors on platelets. Keywords: Platelet signaling, VASP, Flow cytometry, Clopidogrel