Animals (Sep 2024)

Comparison between a Flash Glucose Monitoring System and a Portable Blood Glucose Meter for Monitoring of Cats with Diabetic Ketosis or Ketoacidosis

  • Jennifer S. Eiermann,
  • Katarina Hazuchova,
  • Beatriz Vidondo,
  • Miguel Campos,
  • Simone Schuller,
  • Yi Cui

DOI
https://doi.org/10.3390/ani14192787
Journal volume & issue
Vol. 14, no. 19
p. 2787

Abstract

Read online

Cats with diabetic ketosis or ketoacidosis DK(A) require intensive glucose monitoring. The aim of this study was to assess the agreement between a portable blood glucose meter (PBGM) and a flash glucose monitoring system (FGMS; FreeStyle Libre 2.0 Abbott®) measuring interstitial glucose in cats with DK(A). Ten client-owned cats with naturally occurring DK(A) were prospectively enrolled. Glucose concentrations were assessed with both methods every 1–3 h until resolution of DK(A), and every 4–8 h thereafter. While the median FGMS measured glucose concentration (14.3 mmol/L) was significantly lower than the median PBGM measured glucose concentration (19 mmol/L) (p p 5.5 mmol/L), 81% of results deviated >15% from the PBGM measurements. There was significant inter-individual variation in the difference between glucose concentrations measured by the FGMS and PBGM (p < 0.001). In spite of being more easily tolerated and easier to use, currently this method cannot be recommended for routine monitoring of cats with DK(A).