Química Nova (Dec 2002)

Desenvolvimento de um dispositivo de baixo custo para medidas por quimiluminescência Development of a low cost device for chemiluminescence measurements

  • Eduardo P. Borges,
  • Elizabeth N. Fernandes,
  • Fábio R. P. Rocha,
  • Boaventura F. Reis

DOI
https://doi.org/10.1590/S0100-40422002000700021
Journal volume & issue
Vol. 25, no. 6b
pp. 1191 – 1193

Abstract

Read online

A simple and low cost device (ca. US$ 150) that comprises two photodiodes fixed in lab-made Perspex flow cell is proposed for chemiluminescence measurements. The characteristics of the device (large observation window and reduced thickness) allow maximizing the amount of the emitted radiation detected. A sensitivity improvement of ca. 50 % was observed by employing two photodiodes for signal measurements. The performance of the device was assessed by the oxidation of luminol by hydrogen peroxide, yielding a linear response within the range of 2.50 to 500 µmol L-1 H2O2. The detection limit was estimated as 0.8 µmol L-1 hydrogen peroxide which is comparable with those obtained by using equipments based on photomultipliers.

Keywords