Cardiovascular Ultrasound (Jun 2012)
Can echocardiographic particle image velocimetry correctly detect motion patterns as they occur in blood inside heart chambers? A validation study using moving phantoms
Abstract
Abstract Aims To validate Echo Particle Image Velocimetry (PIV) Methods High fidelity string and rotating phantoms moving with different speed patterns were imaged with different high-end ultrasound systems at varying insonation angles and frame rates. Images were analyzed for velocity and direction and for complex motion patterns of blood flow with dedicated software. Post-processing was done with MATLAB-based tools (Dflow, JUV, University Leuven). Results Velocity estimation was accurate up to a velocity of 42 cm/s (r = 0.99, p Conclusion Echo-PIV appears feasible. Velocity estimates are accurate, but the maximal detectable velocity depends strongly on acquisition parameters. Direction estimation works sufficiently, even at higher velocities. Echo-PIV appears to be a promising technical approach to investigate flow patterns by echocardiography.