Heliyon (Sep 2024)

Spectrochip-based Calibration Curve Modeling (CCM) for Rapid and Accurate Multiple Analytes Quantification in Urinalysis

  • Cheng-Hao Ko,
  • Ashenafi Belihu Tadesse,
  • Abel Chernet Kabiso

Journal volume & issue
Vol. 10, no. 18
p. e37722

Abstract

Read online

Most urine test strips are intended to enable the general population to rapidly and easily diagnose potential renal disorders. It is semi-quantitative in nature, and although the procedure is straightforward, certain factors will affect the judgmental outcomes. This study describes rapid and accurate quantification of twelve urine test strip parameters: leukocytes, nitrite, urobilinogen, protein, pH, occult blood, specific gravity, ketone, bilirubin, glucose, microalbumin, and creatinine using a micro-electromechanical system (MEMS)-based spectrophotometer, known as a spectrochip. For each parameter, absorption spectra were measured three times independently at eight different concentration levels of diluted standard solutions, and the average spectral intensities were calculated to establish the calibration curve under the characteristic wavelength (λc). Then, regression analysis on the calibration curve was performed with GraphPad Prism software, which revealed that the coefficient of determination (R2) of the modeled calibration curves was greater than 0.95. This result illustrates that the measurements exceed standard levels, confirming the importance of a spectrochip for routine multi-parameter urine analysis. Thus, it is possible to obtain the spectral signal strength for each parameter at its characteristic wavelength in order to compare directly with the calibration curves in the future, even in situations when sample concentration is unknown. Additionally, the use of large testing machines can be reduced in terms of cost, time, and space by adopting a micro urine testing platform based on spectrochip, which also improves operational convenience and effectively enables point-of-care (POC) testing in urinalysis.

Keywords