Sensors (Jan 2023)

Comparison of X-ray Radiant Power Absolute Measurement between a Free-Air Ionization Chamber and a Cryogenic Electrical Substitution Radiometer

  • Fan Li,
  • Yidong Zhao,
  • Peiwei Wang,
  • Kun Tang,
  • Lei Zheng

DOI
https://doi.org/10.3390/s23021006
Journal volume & issue
Vol. 23, no. 2
p. 1006

Abstract

Read online

Absolute measurement of radiant power in the X-ray region is essential for many applications in astrophysics, spectroscopy, and X-ray diagnostics. Comparison between different measuring methods is an effective way to check their reliability. In the present work, a comparison of X-ray radiant power absolute measurement between a free-air ionization chamber and a cryogenic electrical substitution radiometer was performed at Beijing Synchrotron Radiation Facility. The absolute radiant power obtained by these two methods were mutually compared via a transfer standard detector’s spectral responsivity at a photon energy of 10 keV. The result of the comparison showed that the difference was 0.47%. A conclusion was reached that the free-air ionization chamber and the cryogenic electrical substitution radiometer agreed within the combined relative uncertainty of 3.35%.