BMC Oral Health (Jul 2022)
Validation of an implant stability measurement device using the percussion response: a clinical research study
Abstract
Abstract Background Several devices have been developed to measure implant-bone stability as an indicator of successful implant treatment; these include Osstell®, which measures the implant stability quotient (ISQ), and the more recent AnyCheck®, which relies on percussion for the implant stability test (IST). These devices make it possible to measure implant stability. However, no studies have compared the performance of AnyCheck® and Osstell® (i.e., IST and ISQ values) in clinical practice. Therefore, this study aimed to determine the correlation between primary and secondary implant stability using the Osstell® and AnyCheck® devices. Methods Ten patients (7 women; age [mean ± standard deviation]: 49.1 ± 13.3 years) with partially edentulous jaws who received a total of 15 implants were included. IST (AnyCheck®) and ISQ (Osstell®) values were measured immediately after implantation and at 1, 2, 3, 4, and 6 weeks post-implantation. Each measurement was performed three times, and the average value was used as the result. The correlation between measurements obtained using the two devices was determined using Spearman's rank correlation coefficient. Results The IST values ranged from 79.1 ± 2.87 to 82.4 ± 2.65. The ISQ values ranged from 76.0 ± 2.8 to 80.2 ± 2.35. Spearman's rank correlation coefficient was r = 0.64 immediately after implantation, r = 0.29 at 1 week, r = 0.68 at 2 weeks, r = 0.53 at 3 weeks, r = 0.68 at 4 weeks, and r = 0.56 at 6 weeks. A positive correlation was found in all cases, except at week 1 when the correlation was weak; the IST and ISQ values decreased the most during the first postoperative week and increased during the second week. The IST values were also slightly higher at all measurement points. Conclusion The ability to assess implant stability without removing the abutment during healing is essential for determining the timing of loading without the risk of bone resorption. The results of this study suggest that AnyCheck® is useful for determining primary and secondary implant stability.
Keywords