Abstract

Read online

MRI-based gel dosimeters are attractive systems for the evaluation of complex dose distributions in radiotherapy. In particular, the nanocomposite Fricke gel dosimeter is one among a few dosimeters capable of accurately evaluating the dose distribution of heavy ion beams. In contrast, reduction of the scanning time is a challenging issue for the acquisition of three-dimensional volume data. In this study, we investigated a three-dimensional dose distribution measurement method for heavy ion beams using variable flip angle (VFA), which is expected to significantly reduce the MRI scanning time. Our findings clarified that the whole three-dimensional dose distribution could be evaluated within the conventional imaging time (20 min) and quality of one cross-section.