Clinical Case Reports (Feb 2020)
Fracture of an epicardial left ventricular lead implanted at open‐heart surgery in anticipation of future need for cardiac resynchronization therapy
Abstract
Abstract Epicardial left ventricular leads can be implanted at open‐heart surgery for cardiac resynchronization therapy. We report a 2‐year‐old fractured epicardial left ventricular lead detected at generator implant. It highlights the importance of good surgical implant technique and of rigorous lead evaluation for signs of impending failure at generator implant.