Sensors (Mar 2018)

Accurate Determination of the Q Quality Factor in Magnetoelastic Resonant Platforms for Advanced Biological Detection

  • Ana Catarina Lopes,
  • Ariane Sagasti,
  • Andoni Lasheras,
  • Virginia Muto,
  • Jon Gutiérrez,
  • Dimitris Kouzoudis,
  • José Manuel Barandiarán

DOI
https://doi.org/10.3390/s18030887
Journal volume & issue
Vol. 18, no. 3
p. 887

Abstract

Read online

The main parameters of magnetoelastic resonators in the detection of chemical (i.e., salts, gases, etc.) or biological (i.e., bacteria, phages, etc.) agents are the sensitivity S (or external agent change magnitude per Hz change in the resonance frequency) and the quality factor Q of the resonance. We present an extensive study on the experimental determination of the Q factor in such magnetoelastic resonant platforms, using three different strategies: (a) analyzing the real and imaginary components of the susceptibility at resonance; (b) numerical fitting of the modulus of the susceptibility; (c) using an exact mathematical expression for the real part of the susceptibility. Q values obtained by the three methods are analyzed and discussed, aiming to establish the most adequate one to accurately determine the quality factor of the magnetoelastic resonance.

Keywords