Indian Heart Journal (Sep 2018)

Intra- and inter-operator reproducibility of automated cloud-based carotid lumen diameter ultrasound measurement

  • Luca Saba,
  • Sumit K. Banchhor,
  • Tadashi Araki,
  • Klaudija Viskovic,
  • Narendra D. Londhe,
  • John R. Laird,
  • Harman S. Suri,
  • Jasjit S. Suri

Journal volume & issue
Vol. 70, no. 5
pp. 649 – 664

Abstract

Read online

Background: Common carotid artery lumen diameter (LD) ultrasound measurement systems are either manual or semi-automated and lack reproducibility and variability studies. This pilot study presents an automated and cloud-based LD measurements software system (AtheroCloud) and evaluates its: (i) intra/inter-operator reproducibility and (ii) intra/inter-observer variability. Methods: 100 patients (83 M, mean age: 68 ± 11 years), IRB approved, consisted of L/R CCA artery (200 ultrasound images), acquired using a 7.5-MHz linear transducer. The intra/inter-operator reproducibility was verified using three operator’s readings. Near-wall and far carotid wall borders were manually traced by two observers for intra/inter-observer variability analysis. Results: The mean coefficient of correlation (CC) for intra- and inter-operator reproducibility between all the three automated reading pairs were: 0.99 (P < 0.0001) and 0.97 (P < 0.0001), respectively. The mean CC for intra- and inter-observer variability between both the manual reading pairs were 0.98 (P < 0.0001) and 0.98 (P < 0.0001), respectively. The Figure-of-Merit between the mean of the three automated readings against the four manuals were 98.32%, 99.50%, 98.94% and 98.49%, respectively. Conclusions: The AtheroCloud LD measurement system showed high intra/inter-operator reproducibility hence can be adapted for vascular screening mode or pharmaceutical clinical trial mode. Keywords: Atherosclerosis, Carotid, Cloud-based, Reproducibility, Reliability