Journal of Cardiovascular Magnetic Resonance (Jul 2011)

Contrast-enhanced CMR in patients after percutaneous closure of the left atrial appendage: A pilot study

  • Petersen Steffen E,
  • Wunderlich Nina,
  • Mohrs Oliver K,
  • Pottmeyer Anselm,
  • Kauczor Hans-Ulrich

DOI
https://doi.org/10.1186/1532-429X-13-33
Journal volume & issue
Vol. 13, no. 1
p. 33

Abstract

Read online

Abstract Background To evaluate the feasibility and value of first-pass contrast-enhanced dynamic and post-contrast 3D CMR in patients after transcatheter occlusion of left atrial appendage (LAA) to identify incorrect placement and persistent leaks. Methods 7 patients with different occluder systems (n = 4 PLAATO; n = 2 Watchman; n = 1 ACP) underwent 2 contrast-enhanced (Gd-DOTA) CMR sequences (2D TrueFISP first-pass perfusion and 3D-TurboFLASH) to assess localization, artifact size and potential leaks of the devices. Perfusion CMR was analyzed visually and semi-quantitatively to identify potential leaks. Results All occluders were positioned within the LAA. The ACP occluder presented the most extensive artifact size. Visual assessment revealed a residual perfusion of the LAA apex in 4 cases using first-pass perfusion and 3D-TurboFLASH indicating a suboptimal LAA occlusion. By assessing signal-to-time-curves the cases with a visually detected leak showed a 9-fold higher signal-peak in the LAA apex (567 ± 120% increase from baseline signal) than those without a leak (61 ± 22%; p Conclusion This CMR pilot study provides valuable non-invasive information in patients after transcatheter occlusion of the LAA to identify correct placement and potential leaks. We recommend incorporating CMR in future clinical studies to evaluate new device types.