Journal of International Medical Research (Nov 2021)
Enumeration of viable CD34+ cells in cord blood using a novel stem cell enumeration kit
Abstract
Objective To assess the detection performance of the hematopoietic stem cell enumeration kit developed by BD Biosciences. Methods Cord blood samples were prepared using a hematopoietic stem cell enumeration kit developed by BD Biosciences and Stem-Kit reagents from Beckman Coulter. CD34+ cells were enumerated using a BD FACSCanto instrument and FACSDiva software. Results A total of 519 samples were analyzed in this study. The hematopoietic stem cell enumeration kit developed by BD Biosciences yielded absolute counts of CD34-positive cells that were on average 8.7% lower than Beckman Coulter Stem-Kit reagents (range: −5.7% to−14.7%). The BD Biosciences kit yielded relative counts that were on average 9.9% higher compared with Beckman Coulter Stem-Kit reagents (range: −2.1% to +13.8%). The intraclass correlation coefficients for absolute and relative counts of CD34-positive cells were 0.9967 (95% confidence interval [CI]: 0.9961–0.9972) and 0.9512 (95% CI: 0.9423–0.9587) for the BD Biosciences and Beckman Coulter kits, respectively. Conclusions The hematopoietic stem cell enumeration kit developed by BD Biosciences can be used to enumerate CD34-positive stem cells from cord blood samples.