Indian Journal of Ophthalmology (Jan 2021)

Assessment of precision of astigmatism measurements taken by a sweptsource optical coherence tomography biometer - IOLMaster 700

  • Ajay Sharma,
  • Akanksha Batra

DOI
https://doi.org/10.4103/ijo.IJO_2776_20
Journal volume & issue
Vol. 69, no. 7
pp. 1760 – 1765

Abstract

Read online

Purpose: To assess the repeatability of ocular measurements, particularly astigmatism, taken using the latest version of an swept-source optical coherence tomography (SS-OCT) device, IOLMaster 700. Methods: This prospective observational study done in a private eye care centre. Study included 213 eyes of 152 patients diagnosed with cataract. Axial length (AL), Anterior corneal astigmatism (ΔK), Total corneal astigmatism (ΔTK), Aqueous depth (AD), Lens thickness (LT), Iris barycentre distance (IBD) and White to white distance (WTW) were analysed; three measurements were taken for each patient in the same sitting by the same examiner in an undilated state using the SS-OCT device. Repeatability was analysed using intraclass coefficient (ICC) and coefficient of variation (CV) measures, and astigmatism values were transformed into vector components (J0/J45) and categorised based on magnitude. Results: Astigmatism vector components (J0/J45) showed poor repeatability (ICC 0), while iris coordinates (IBD) and astigmatism magnitude had good correlation (ICC > 0.9) but more variations between the readings which was least for magnitude ≥ 1.5D; however, rest of the parameters demonstrated excellent repeatability (ICC > 0.9 with P value < 0.05). Conclusion: IOLMaster 700 demonstrated good reliability for the parameters measured, although, for astigmatism and iris barycentre distance, more number of readings and cross reference with other devices may be required to get clinically accurate results.

Keywords