Indian Journal of Ophthalmology (Jan 2021)

Comparison of central corneal thickness using non-contact tono-pachymeter (Tonopachy) with ultrasound pachymetry in normal children and in children with refractive error

  • Kiruthikadevi Thiagarajan,
  • Kavitha Srinivasan,
  • Keerthi Gayam,
  • Venkatesh Rengaraj

DOI
https://doi.org/10.4103/ijo.IJO_364_21
Journal volume & issue
Vol. 69, no. 8
pp. 2053 – 2059

Abstract

Read online

Purpose: To compare the central corneal thickness (CCT) measured by non-contact tono-pachymeter [Tonopachy (TP)] with the gold standard ultrasound pachymetry (UP) in normal children and in children with refractive error. Methods: This cross-sectional study involved 95 normal children (190 eyes) and 123 children with refractive error (246 eyes), a total of 218 children (436 eyes) aged 7-15 years. After refraction and complete ophthalmic evaluation, axial length was measured with IOLMaster 700, CCT was measured with TP followed by UP. The correlation between the CCT measurements obtained with the two methods was analysed by Intraclass correlation coefficient (ICC) and the limits of agreement were assessed with Bland-Altman analysis plot. Results: Mean CCT measured with TP was 537.46 ± 36.41 μm and by UP was 537.17 ± 39.80 μm in normal children (P = 0.79) and in refractive error group, the mean CCT was 533.50 ± 34.91 μm by TP and 531.60 ± 36.30 μm by UP (P = 0.04). The 95% limits of agreement between TP and UP ranged from -19.2 to + 21.6 μm (mean = 1.20) for total children, -20.8 to + 21.4 μm (mean = 0.29) for normal group and -18.0 to + 21.8 μm (mean = 1.90) for refractive error group. ICC for CCT measurement between TP and UP was 0.980 for total children and refractive error group and 0.981 for normal group. Conclusion: The CCT measurement obtained by TP showed an excellent agreement with UP. Hence non-contact TP can be used to assess CCT in children above seven years of age.

Keywords