Journal of Isotopes (Oct 2021)

Quantitative Assessment of the Dual-Probe WIZARD2 2470 Gamma Counter

  • ZHAO Fukuan;XU Xijie;WANG Yan

DOI
https://doi.org/10.7538/tws.2021.34.03.0282
Journal volume & issue
Vol. 34, no. 5

Abstract

Read online

Gamma counter is one of the most important instruments for precise quantitative assessment of the radioactivity. The accuracy of instrument performance and standardization of experimental operation are closely related to the reliability and authenticity of experimental results. In this study, three positron nuclides 18F, 68Ga and 89Zr were used to systematically evaluate the dual-probe detection efficiency, detection limit, stability and linear range on the dual-probe WIZARD2 2470 gamma counter. The results suggested that the two probes have the same detection efficiency, and the influence of the environmental background should be considered during the experiment. Good linearity relationship appeared between gamma count and radioactivity in the range of 1×103-6×106 min-1, with excellent quantitative accuracy and stability. The quantitative method of the dual-probe WIZARD2 2470 gamma counter in this study is expected to provide a reference for the standardized use of similar instruments in other laboratories.

Keywords