Radiation Oncology (Jun 2019)
Circular collimator arc versus dynamic conformal arc treatment planning for linac-based stereotactic radiosurgery of an intracranial small single lesion: a perspective of lesion asymmetry
Abstract
Abstract Background Although circular collimator arcs (CCA) and dynamic conformal arcs (DCA) are commonly used linear accelerator-based treatment planning techniques for intracranial stereotactic radiosurgery (SRS) of a small single lesion, these two techniques have not been rigorously compared in terms of tumor shape. Therefore, this study compared clinical CCA plans with re-planned DCA plans using conformity index (CI) and V12Gy (volume of normal brain tissue receiving 12 Gy or higher) from a perspective of asymmetry (Asym) of planning target volume (PTV). Methods Ninety-five clinical CCA plans delivered for a small single lesion with PTV size 6.12%. Conclusions The DCA technique is recommended when a lesion has PTV Asym > 6.12% regardless of PTV size. For lesions with PTV Asym ≤6.12%, a technique choice would depend on the preference of CI or V12Gy.
Keywords