Journal of Cardiovascular Development and Disease (Apr 2023)
Radiation Exposure Reduction by Digital Variance Angiography in Lower Limb Angiography: A Randomized Controlled Trial
Abstract
Background: digital variance angiography (DVA) provides higher image quality than digital subtraction angiography (DSA). This study investigates whether the quality reserve of DVA allows for radiation dose reduction during lower limb angiography (LLA), and compares the performance of two DVA algorithms. Methods: this prospective block-randomized controlled study enrolled 114 peripheral arterial disease patients undergoing LLA into normal dose (ND, 1.2 µGy/frame, n = 57) or low-dose (LD, 0.36 µGy/frame, n = 57) groups. DSA images were generated in both groups, DVA1 and DVA2 images were generated in the LD group. Total and DSA-related radiation dose area product (DAP) were analyzed. Image quality was assessed on a 5-grade Likert scale by six readers. Results: the total and DSA-related DAP were reduced by 38% and 61% in the LD group. The overall visual evaluation scores (median (IQR)) of LD-DSA (3.50 (1.17)) were significantly lower than the ND-DSA scores (3.83 (1.00), p p p < 0.001). Conclusions: DVA significantly reduced the total and DSA-related radiation dose in LLA, without affecting the image quality. LD-DVA2 images outperformed LD-DVA1, therefore DVA2 might be especially beneficial in lower limb interventions.
Keywords