PLoS ONE (Jan 2012)
Point-of-care test for detection of urogenital chlamydia in women shows low sensitivity. A performance evaluation study in two clinics in Suriname.
Abstract
BackgroundIn general, point-of-care (POC) tests for Chlamydia trachomatis (Ct) show disappointing test performance, especially disappointing sensitivity results. However, one study sponsored by the manufacturer (Diagnostics for the Real World) reported over 80% sensitivity with their Chlamydia Rapid Test (CRT). We evaluated the performance of this CRT in a non-manufacturer-sponsored trial.MethodsBetween July 2009 and February 2010, we included samples from 912 women in both high- and low-risk clinics for sexually transmitted infections (STIs) in Paramaribo, Suriname. Sensitivity, specificity, positive- and negative predictive values (PPV and NPV) for CRT compared to NAAT (Aptima, Gen-Probe) were determined. Quantitative Ct load and human cell load were determined in all CRT and/or NAAT positive samples.ResultsCRT compared to NAAT showed a sensitivity and specificity of 41.2% (95% CI, 31.9%-50.9%) and 96.4% (95% CI, 95.0%-97.5%), respectively. PPV and NPV were 59.2% (95% CI, 47.5%-70.1%) and 92.9% (95% CI, 91.0%-94.5%), respectively. Quantitative Ct bacterial load was 73 times higher in NAAT-positive/CRT-positive samples compared to NAAT-positive/CRT-negative samples (pConclusionsThe sensitivity of CRT for detecting urogenital Ct in this non-manufacturer-sponsored study did not meet the expectations as described previously. The CRT missed samples with a low Ct load. Improved POC are needed as meaningful diagnostic to reduce the disease burden of Ct.