Scientific Reports (Jul 2024)

Wear and corrosion of titanium alloy spinal implants in vivo

  • Hangyu Ji,
  • Xinhui Xie,
  • Zhe Jiang,
  • Xiaotao Wu

DOI
https://doi.org/10.1038/s41598-024-68057-8
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 9

Abstract

Read online

Abstract To investigate the wear and corrosion of titanium alloy spinal implants in vivo, we evaluated removed implants and their surrounding scar tissues from 27 patients between May 2019 and April 2021. We performed scanning electron microscopy, energy-dispersive X-ray spectroscopy, and histological analysis. The results revealed metal-like particles in the soft tissues of seven patients, without any considerable increase in inflammatory cell infiltration. Patients with fractures showed lower percentages of wear and corrosion compared with other patients (42% and 17% vs. 59% and 26%). Polyaxial screws exhibited higher wear and corrosion percentages (53% and 23%) compared with uniaxial screws (39% and 3%), although in patients with fracture, the reverse was observed (20% and 0% vs. 39% and 3%). We found that titanium alloy spinal implants experience some degree of wear and corrosion in vivo. The titanium alloy particles formed by wear exhibited good histocompatibility, not causing inflammation, foreign body reactions, or osteolysis. Therefore, spinal implants should be removed cautiously when treating titanium alloy spinal metallosis. The wear and corrosion of the implants increase with the increase in implantation time, although the screw structure does not significantly affect these changes.