Scientific Reports (Aug 2022)
Deviation from the planned axis of three toric intraocular lenses
Abstract
Abstract In this study, we retrospectively evaluated the deviation from the planned axis of 3 Toric intraocular lenses (TIOL). Included in the study 190 eyes, operated by two surgeons using two different manual marking techniques. The patients were implanted with either AcrySof IQ Toric SN6AT (Alcon) (n = 90), POD FT (PhysIOL) (n = 50), or TECNIS Symfony Toric (J&J) (n = 50). At least 1 month postoperatively, the IOL was photographed, and the axis was measured using a designed software. The difference between the planned and actual axis was defined as axis deviation. The effect of IOL type, astigmatism direction, and marking techniques on the average degree and direction of the IOL deviation were evaluated and compared. There was no significant difference in the average deviation between the IOLs (TECNIS Symfony: 4.03° ± 4.34, POD FT: 3.52° ± 3.38, and SN6AT: 4.24° ± 4.10), and its direction (55.8%, 39.0%, and 56.6% clockwise (CW) deviation, respectively). With the rule, astigmatism had significantly more CW deviation compared with against the rule and oblique astigmatism (64.3%, 43.8%, and 41.7%, respectively, P = 0.027), but the average deviation was similar. The marking techniques did not influence the degree or direction of the deviation.