Practical Laboratory Medicine (Mar 2024)

Biological variation of PIVKA-II in blood serum of healthy subjects measured by automated electrochemiluminescent assay

  • Antonín Jabor,
  • Zdenek Kubíček,
  • Jitka Čásenská,
  • Tereza Vacková,
  • Vanda Filová,
  • Janka Franeková

Journal volume & issue
Vol. 39
p. e00389

Abstract

Read online

Background: Prothrombin/Protein Induced by Vitamin K Absence-II (PIVKA-II) is a candidate biomarker of hepatocellular cancer, recommended both for diagnostics and monitoring. The aim was to evaluate biological variation (BV) of serum PIVKA-II. Methods: Within-subject (CVI) and between-subject (CVG) BV estimates were assessed in 14 healthy volunteers in a 6-week protocol. Serum concentrations of PIVKA-II were measured by a Roche Elecsys PIVKA-II diagnostic kit (cobas e8000). Precision (CVA) was assessed from duplicate measurements of all volunteers' samples. Two methods were used for the estimation of CVI: SD-ANOVA and CV-ANOVA method. We calculated the index of individuality (II) and reference change value. The experiment was fully compliant with EFLM database checklist. Results: The CVI of PIVKA-II in healthy persons, as calculated by two statistical methods, were 8.2% (SD-ANOVA with CVA of 3.2%) and 9.4% (CV-ANOVA) with CVA of 2.7%). The CVG was 19.5% (SD-ANOVA), and respective II and RCV were 0.42 and 24.4%. Conclusions: CVI and CVG of PIVKA-II were 8.2% and 19.5%, respectively, with CVA below 4%. The low II and RCV below 25% enable the use of this biomarker both for diagnostics and monitoring. More data are needed before the introduction of PIVKA-II into clinical practice.