Data in Brief (Dec 2020)

Uncertainty analysis of a non-automatic weighing instrument from calibration data on scales according to the SIM guide

  • Germán Roca-Gómez,
  • Ulises Ospino-López,
  • Cristian Antonio Pedraza-Yepes,
  • Oscar Fabián Higuera-Cobos,
  • José Daniel Hernández-Vásquez

Journal volume & issue
Vol. 33
p. 106436

Abstract

Read online

This research was motivated by technical-economic challenges imposed by mass metrology, specifically, in matters concerning calibration methods of non-automatic type weighing instruments (i.e.: digital scales). In order to contextualize the problem detected, in the industry there are different processes of mass measurement that are controlled by digital scales, such as: mass of liquids, chemicals, food, body mass of a person. In these processes, the scale is used in the following four conditions for mass measurement: (i) ascending and descending load, returned to zero; (ii) ascending and descending load, without the need to return to zero; (iii) only with ascending load and (iv) only with descending load. In this context and, maintaining the principles for the calibration of a measurement instrument in which it must be carried out under the same operating conditions as the instrument, metrology laboratories must knowing the metrological reliability (i.e.: errors and uncertainties) for each situation. This is exactly the main motivation for the development of the research. Thus, the experimental data obtained in a research laboratory under controlled environmental conditions allowed obtaining a minimum expanded uncertainty associated with the mass measurement of 0.0012 kg (k=2; confidence level: 95%).