Scientific Reports (Jul 2024)

Multicenter validation study for automated left ventricular ejection fraction assessment using a handheld ultrasound with artificial intelligence

  • Nobuyuki Kagiyama,
  • Yukio Abe,
  • Kenya Kusunose,
  • Nahoko Kato,
  • Tomohiro Kaneko,
  • Azusa Murata,
  • Mitsuhiko Ota,
  • Kentaro Shibayama,
  • Masaki Izumo,
  • Hiroyuki Watanabe

DOI
https://doi.org/10.1038/s41598-024-65557-5
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 8

Abstract

Read online

Abstract We sought to validate the ability of a novel handheld ultrasound device with an artificial intelligence program (AI-POCUS) that automatically assesses left ventricular ejection fraction (LVEF). AI-POCUS was used to prospectively scan 200 patients in two Japanese hospitals. Automatic LVEF by AI-POCUS was compared to the standard biplane disk method using high-end ultrasound machines. After excluding 18 patients due to infeasible images for AI-POCUS, 182 patients (63 ± 15 years old, 21% female) were analyzed. The intraclass correlation coefficient (ICC) between the LVEF by AI-POCUS and the standard methods was good (0.81, p 0.80), AI-POCUS tended to underestimate LV volumes for larger LV (overall bias 42.1 mL for end-diastolic volume). These trends were mitigated with a newer version of the software tuned using increased data involving larger LVs, showing similar correlations (ICC > 0.85). In this real-world multicenter study, AI-POCUS showed accurate LVEF assessment, but careful attention might be necessary for volume assessment. The newer version, trained with larger and more heterogeneous data, demonstrated improved performance, underscoring the importance of big data accumulation in the field.