BMC Medical Imaging (Jan 2023)

The comparison of diffusion tensor imaging in human hearts between 1.5 T and 3.0 T

  • Xiaodan Li,
  • Rui Chen,
  • Xi Xu,
  • Zebin Xiao,
  • Xiaoyu Wei,
  • Yuelong Yang,
  • Zhongping Zhang,
  • Zhigang Wu,
  • Yanjie Zhu,
  • Hui Liu

DOI
https://doi.org/10.1186/s12880-023-00969-9
Journal volume & issue
Vol. 23, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Background The aim was to compare the diffusion tensor imaging (DTI) indices derived from human hearts between 1.5 T and 3.0 T scanners. Additionally, the reproducibility of DTI indices was assessed between 1.5 T and 3.0 T scanners. Methods A total of 18 ex-vivo hearts were derived from patients who underwent heart transplantation. The DTI schemes were performed at 1.5 T and 3.0 T, respectively. Then, the same slices from each ex-vivo heart were selected for image analysis. The student’s t-test or Wilcoxon-rank test was used to compare the statistical differences. The agreement of DTI indices was mainly reported as the interclass correlation coefficient (ICC). Results No significant differences (all P > 0.05) were found in the DTI indices between 1.5 T and 3.0 T scanners. Interestingly, the ICC of all DTI indices was relatively lower with a low b-value. The reproducibility of the helix angle (HA) was relatively lower when compared to the other DTI indices. Conclusion The DTI indices of ex-vivo human hearts between 1.5 T and 3.0 T scanners had no significant differences. The consistency of DTI indices needed caution using a low b-value with different field strengths, and the relatively low reproducibility of HA should be considered.